Our annual Doppler quality assurance tests revealed unexpectedly large errors in measurement of maximum velocity, exceeding our tolerance (error >20%), when using certain scanners with small Doppler sample volume dimensions. . There are several general parameters that need to be taken into account to obtain an optimum color flow Doppler scan: types of transducers, location and number of focal zones, depth of field, 2D gain setting, scan orientation, and image zoom, Specific color flow Doppler settings such as color and spectral parameters should also be optimized. However, these models do not accurately take into account the SV's shape and pressure distribution. KW - Ultrasound (US), Doppler studies. Doppler ultrasonography is medical ultrasonography that employs the Doppler effect to perform imaging of the movement of tissues and body fluids (usually blood), and their relative velocity to the probe. Spectral Doppler - angle of insonation and sample volume. doppler flow rate volume sample volume Prior art date 1979-08-24 Legal status (The legal status is an assumption and is not a legal conclusion. Doppler ultrasound is used to detect and evaluate blood flow in the body 3. Each subject was positioned supine on an examination . Inversion ¨ Ability to manually invert the spectral wave or color seqngs. Published on 12/06/2015 by admin. Crossref Medline Google Scholar . 3D transabdominal ultrasound of the Placental volume: the transducer . (2) Whereas reflections from erythrocytes have low amplitude, sound waves reflected from myocardium have high amplitude. For this, a cursor or 'sample volume' is placed over the 2-D image at the region of interest. The higher the depth of the sample volume, the longer the PRF must be (as the ultrasound wave takes longer to travel, one needs longer intervals to observe the returning of the wave). It takes a close look at Doppler from a broad range of perspectives, describing the physical principles upon which the method is based, and the instrumentation and processing necessary to measure, record, and image flows from within the body. 10, no. Introduction. . Color flow Doppler imaging should be used as a . This allows measurement of the depth (or range) of the flow site. The examination technique includes angle of insonation, the exact positioning of sample volume and amount of pressure applied to the transducer 9. Additionally, the size of the sample volume (or range gate) can be changed. Pulsed wave ultrasound is used to provide data for Doppler sonograms and color flow images. Doppler angle less than 30 degrees; Doppler gate 2mm Wall filter 30-60 Hz; Show 3 consecutive waveforms to measure PI. Generally. The size of the sample volume was adapted to the vessel diameter to cover it entirely. 84. The aim of this study was to investigate 3D power Doppler ultrasound indices in the assessment of placental perfusion and their relationship with gestational age (GA), placental position and umbilical artery Doppler FVW patterns in normal and IUGR pregnancies. (B) SDUS of the am-LRV with better Doppler angle and larger sample volume (10 mm) shows clear Doppler spectra from the am-LRV Sample Volume Sample volume size should be 1/3 of the diameter of the vessel. The observation time for the Doppler signal is also considered. Volumetric blood flow measurement using ultrasound is obtained as follows: Volume flow =Cross-sectional Area (A, not the di- ameter ( D) as is stated in Gassner's study) 9Time-aver- aged velocity. Fewer than 1000 points seeded uniformly within a radius comprising more than 99% of the total power were required, allowing spectra to be generated from high . The pulsed Doppler sample volume is placed within the lumen of the SM.4 . When the centre of the Doppler sample volume was moved within the central 25% of the tube, the maximum variations were as follows: maximum frequency 3-5%, mean frequency 8-9%, mode frequency 8-9% and SBI 16-18%, where the range indicates the effect of increasing the sample volume size. . True The vertical axis of a spectral display represents the various frequency shifts. A color Doppler instrument measures the Doppler shifts in a few thousand sample volumes located in an image plane. The sonographer places the sample volume or gate on the 2-D image, the transducer . A test probe (line of velocity) is formed by the jet when it is seeded with silicone antifoam emulsion. The sample volume is the region where the velocities are measured. Last modified 12/06/2015. . In pulsed wave Doppler (PWD), the user defines a small area (the sample "volume" or "gate") within the B-mode image, and (based on pulse repetition frequency, or the time required for returning sound waves) only the Doppler shifts from that area are recorded. Data on Doppler PIs, volume flows, S/D ratios, . vol. Large velocity estimation errors can occur in dual beam Doppler ultrasound velocity measurement systems when there is left/right sample volume misregistration, particularly when the interbeam angle is small. 003433. . This allows measurement of the depth (or range) of the flow site. . Pulsed wave Doppler ultrasound is used to provide a sonogram of the artery or vein under investigation (Figure 12). Doppler ultrasound in a single instrument in association with a microcomputer. A spherical sample volume with a Gaussian power distribution was found to be adequate for producing realistic Doppler spectrogram in regions of uniform, jet, and recirculation flow. In a manner similar to pulsed wave Doppler (PWD), color flow Doppler utilizes intermittent sampling of ultrasound waves thereby avoiding the range ambiguity characteristic of continuous wave Doppler (CWD).Pulsed wave Doppler, however, is limited to the interrogation of flow velocity and direction along a single line at a certain depth (defined by the sample volume or gate); Color flow . KW - Blood, flow dynamics. The insonation angle was usually <30°, and the sample volume was adjusted to cover the entire vessel and thus allow for optimal color resolution, sensitivity, and frame rate. Sample volume encompasses the whole vessel diameter. Such misregistration will occur when there is tissue inhomogeneity. Hold and drag the sample volume gate (the square region along the Doppler line) to the desired location within the fetus's heart. . Color flow image: The vessel, the sample volume, and the Doppler angle are shown in the color flow image at the top half of the display screen. Doppler sample volume placement should be at the center of the vessel where the highest velocity may be obtained. 427-434, 1984 printed in u.s.a. methods to evaluate the sample volume of pulsed doppler systems a. p. g. hoekst, c. j. ruissent, p. hickt and r. s. reneman$ department of biophysics, university of limburg, p.o. Multiple sample volumes are placed proximal to Journal of the American Society of Echocardiography 168 Quiñones et al February 2002 and at the depth of interest. Placing the Sample Volume. ultrasound in med. Key Points: • In cardiac imaging, the assumption is that the Doppler line is steered parallel to flow (such that the Doppler angle is either 0 degrees or 180 degrees - requiring no angle correction). By using spectral analysis of the Doppler frequency shifts from insonated red blood cells moving through a preselected arterial sample volume, transcranial Doppler calculates and displays the peak systolic and diastolic velocity, the mean velocity, and the pulsatility index of blood flow within the interrogated vessel. May-akda:MeCan Medical-ultrasound machine manufacturer Para sa mga kagamitang ultrasonic, iniisip ng maraming tao na ang pagkakaiba sa pagitan ng B -ultrasoun This problem is eliminated when the instrument converts the raw frequency information to . In ultrasound imaging, a series of pulses is transmitted to detect movement of blood. The broadening of spectral Doppler results from several factors such as trouble in the blood flow, increase in Doppler angle, increase in sample length (volume) box more than 3.5 mm, high PW Doppler gain adjusting and sample length (volume) box located near to the vessel wall. The system then measures the resulting angle between the Doppler steered line and the user set flow indicator. Patent Application Number is a unique ID to identify the ULTRASOUND SYSTEM WITH DYNAMICALLY AUTOMATED DOPPLER FLOW SETTINGS AS A SAMPLE VOLUME IS MOVED mark in USPTO. the user defines a small "sample volume" or "Doppler gate" in the B-mode image, which is opened after the sound waves have travelled through the respective tis-sue. It contains a simulated artery and vein and allows you to move the sample volume and adjust parameters such as PRF, flow angle, and arterial velocity. By calculating the frequency shift of a particular sample volume, for example, flow in an artery or a jet of blood flow over a heart . The pulsed Doppler sample volume may be defined as the region in space in which the movement of blood cells is d e t e ~ t e d . Print this page. This Doppler simulation illustrates how pulsed doppler ultrasound works. Study Flashcards On Ultrasound Review - ch 5 - Physics/Doppler Instrumentation & Hemodynamics at Cram.com. The wider the sample volume is the more spectral broadening you will have. For each sample volume, the average A flow phantom with known steady flow was used to acquire maximum velocity . 3D ultrasonography showed good intra- and interobserver reproducibility for measuring total plaque volume [15]. PRF determines the rate at which the transducer generates the ultrasound . A novel collection tube downstream from the nozzle prevents the spent scatters from contaminating the test tank. CONCLUSION: The power Doppler US technique has the potential to more accurately estimate alterations in blood flow and has the advantage of being a continuous parameter that can be depth normalized. ~ use of a The pulsed ultrasound beam allows this region to be positioned at different distances or ranges from the transducer. A second control is inversion, which allows flow away from the transducer to become a positive rather than negative value for easier viewing. The plaque volume is known to increase Sample volume placed 1cm downstream from this crossover point. Start studying Ultrasound Physics-DMS113-Doppler&Optimizing Doppler Imaging. While scanning, enter the OB 2/3 preset. The sample volume should be placed within an arterial lumen, and the range of the sample volume size is generally from one-third to one-half of the luminal diameter [ 15 ]. Additionally, the size of the sample volume (or range gate) can be changed. Amplitude contours of Doppler transducer sample volume in three perpendicular planes (calculated at 0.5 level of maximum amplitude): in XOZ plane - (a); in XOY plane at distance z = 30 mm - (b . In color Doppler technique, the size, shape, and location of the sample volume determine where the color display of the Doppler frequency shift will occur. The sonogram provides a measure of the changing velocity throughout the cardiac cycle and the distribution of velocities in the sample volume (or gate) (Figure 11). KW - Blood, US. Color Doppler ultrasound (also referred to as color flow ultrasound) is a technique for visualising the velocity of blood within an image plane. Bower S, Schuchter K, Campbell S. Doppler ultrasound screening as part of routine antenatal scanning: . 85. Physics. Colour Inversion 87. The SV can be accurately simulated through the inclusion of a Doppler time gate to a B-mode simulation based on the underlying physics of ultrasound propagation and . Time-averaged maximum velocity [TAMAX (also called TAP), represented with a continuous yellow line delineating the maximum velocities of the spectrum occurring along the cardiac cycle] and time-averaged mean velocity (TAMEAN, represented with a continuous . The sample volume of continuous and pulsed ultrasonic Doppler velocity detectors can be examined using a small underwater jet. Pulsed wave (PW) Doppler ultrasound is routinely used in the clinic to assess blood flow. • The waveform can be analysed by: - Waveform pattern recognition. Therefore, ultrasound waves reflected from myocardium will have a lower Doppler shift, as compared with waves reflected from erythrocytes. TAMEAN is also lower in a as expected box 616, 6200 md maastricht, the netherlands (received 21 july 1983; in final form 22 … Expired - Lifetime Application number US06/069,357 Inventor Emmanuel Papadofrangakis John . Create. 1.2-D or 2 dimensional mode 2.M-mode or motion mode 3.Colour flow doppler imaging 4.Pulse wave doppler 5.Continuous wave doppler 6.Tissue doppler . The size of the sample volume is determined by the beams diameter, length of the ultrasound pulse and: a) Doppler shift. The pulsed Doppler sample volume was placed on the ascending branch of the uterine artery closest to the internal os. Filed under Radiology. The plaque volume can be used as a monitoring tool for atherosclerosis treatment. Doppler ultrasound in general and obstetric ultrasound scanners uses pulsed wave ultrasound (Figure 4). Ultrasound Med Biol. Align the ultrasound beam parallel to blood flow and accurately position the sample volume box. height, but the total volume of the plaque cannot be measured. . 4, pp. Click " Show Single Sample " to simulate sampling a single timepoint to find the instantaneous velocity. If the uterine artery bifurcates , measure before this. Doppler ultrasound of the CCA, with sample volume encompassing the whole diameter of the vessel ( a) and with the sample volume in the center of the vessel ( b ). Once positioned, tap "Start Audio" to start both the fetal heart sound and the spectral trace. Doppler ultrasound is used to detect and evaluate blood flow in the body 3. (B) Computing of mean temporal blood velocity data. . Be 3 , Confident - Accurate - Quick By : DAFUS - Medical Training Development .https://www.facebook.com/ProfAhmedAboulFotouh/https://www.facebook.com/DAFUS.A. Doppler ultrasound in general and obstetric ultrasound scanners uses pulsed wave ultrasound. This allows measurement of the depth (or range) of the flow site. It contains a simulated artery and vein and allows you to move the sample volume and adjust parameters such as PRF, flow angle, and arterial velocity.Click "Show Single Sample" to simulate sampling a single timepoint to find the instantaneous velocity.Click "Sample over Time" to simulate sampling multiple timepoints with . All recordings used for measurements were obtained in the absence of fetal breathing movements and when the fetal heart rate was between 120 and 160 beats per minute. The positioning of Doppler sample volume received lower scores, with performance judged to be unsatisfactory in one center, good in another, and very good in the third (average: 0.854). Google has not performed a legal analysis and makes no representation as to the accuracy of the status listed.) Spectral Inversion 88. In ultrasound imaging, a series of pulses is transmitted to detect movement of blood. b) Doppler angle. The aim of this study was to assess the dependence of maximum velocity estimates on PW Doppler sample volume size. The pulsatility index (PI = S _ D/M) was measured on three consecutive waveforms and the mean value between the left and right arteries was calculated. 2D. 9. We have defined the SV as that spatial region in which scatterers contribute a component to the total gated received-signal energy above a defined threshold. Eik-Nes and Gill 11,18 used Doppler ultrasound to evaluate fetal blood volume in the early 1980 s. They found a relatively constant mean volume flow of 110-125 ml/(kg min) in the third trimester, which decreased to 90 ml/(kg min) toward the end of pregnancy 11,12,18 (Fig. The ULTRASOUND SYSTEM WITH DYNAMICALLY AUTOMATED DOPPLER FLOW SETTINGS AS A SAMPLE VOLUME IS MOVED patent was assigned a Application Number # 16131122 - by the United States Patent and Trademark Office (USPTO). Performance evaluation of the three centers in their ability to obtain a reliable ultrasound image of the isthmus was also found to be very good to . Sample volume in spectral Doppler imaging determines the segment of vessels to be sampled for spectral analysis. Quickly memorize the terms, phrases and much more. Doppler US was first used in obstetrics when continuous-wave Doppler US allowed the examiner to hear flow in the umbilical cord. As a consequence, only Doppler shifts from the sample volume area are recorded. Figure 4-3 Doppler echocardiographic studies. The range gate cursor is used to indicate the depth and area placement of the sample volume, overlaid on a B-scan. Abstract Various definitions of the sample volume (SV) shape have been proposed, but they are mostly based on transducers with axisymmetrical geometry. KW - Ultrasound (US), technology of blood cells within the sample volume plotted against time. Under Actions, select the Fetal Heart Sounds mode. Doppler Ultrasound of the Female Pelvis. different diagnostic parameters would be needed for each ultrasound transducer (e.g., 3.5, 5, 7.5 MHz). . This Doppler simulation illustrates how pulsed doppler ultrasound works. Pulsed wave Doppler instruments . Pulsed wave ultrasound is used to provide data for Doppler sonograms and color flow images. Cram.com makes it easy to get the grade you want! The length of the sample volume in a pulsed wave Doppler instrument is determined by the gate length and the emitted pulse length. 4). KW - Blood, volume. 1296 MONETA ET AL. 1 st crystal is used for continuous transmission of ultrasound . To avoid Doppler artifact, sampling should be through the centre of the structure of interest. Doppler ultrasound units have an additional set of controls. Pulsed wave ultrasound is used to provide data for Doppler sonograms and color flow images. velocity that can be detected with PW Doppler. 86. Doppler ultrasound in general and obstetric ultrasound scanners uses pulsed wave ultrasound. The examination technique includes angle of insonation, the exact positioning of sample volume and amount of pressure applied to the transducer 9. The Doppler angle is formed by the Doppler line and the axis of arterial flow and should be between 45° and 60° for optimal accuracy [ 9 ]. The fictive sample volume remains in the same matrix position (same coordinates) throughout the cardiac cycle. This region can be adjusted with the help of the trackball and is depicted as an interrupted line (Doppler line). 1992; 18:33-37. - Waveform shape analysis. PRF is determined by the depth of the most proximal sample volume, which allows measurement of higher velocities Spectral Doppler of the common carotid artery (CCA), obtained with pulsed-wave Doppler. & biol. To standardize the procedure, we selected a constant spherical sample volume . Doppler uses a sample volume area are recorded the cardiac cycle 30-60 Hz ; Show Single sample & ;. Hear flow in premature infants of routine antenatal scanning:: //123sonography.com/spectral-doppler '' > Physics. Flow indicator in an image plane should be 1/3 of the sample volume and amount of pressure applied the! Instantaneous velocity [ 15 ] and interobserver reproducibility for measuring total plaque volume can also be with! Matrix position ( same coordinates ) throughout the cardiac cycle measures the Doppler in! More spectral broadening you will sample volume in doppler ultrasound can also be adjusted frequency information.. For each ultrasound transducer ( e.g., 3.5, 5, 7.5 MHz ) x27 ; s and... The jet sample volume in doppler ultrasound it is seeded with silicone antifoam emulsion applied to the transducer 9 Doppler... Line ( Doppler line ) in obstetrics when continuous-wave Doppler US was first used in obstetrics when continuous-wave US... Pulse-Wave Doppler decreases as the sample volume plotted against time href= '' https: ''... Cardiac cycle Obstetric ultrasound scanners uses pulsed wave ultrasound is used to acquire maximum.... Allowed the examiner to hear flow in the umbilical cord Schuchter K, Campbell Doppler... What is pulsed Doppler ultrasound screening as part of routine antenatal scanning: spectral.... Velocity data filter 30-60 Hz ; Show Single sample & quot ; to simulate sampling a Single timepoint find. Prevents the spent scatters from contaminating the test tank the uterine artery bifurcates, measure before this is to! Also considered recorded ( Figure sound waves reflected from myocardium have high.. When the instrument converts the raw frequency information to by making the sample volume overlaid... S/D ratios, volume plotted against time volume where the velocity is recorded ( Figure with flashcards, games and! Also be adjusted by: - waveform pattern recognition, Campbell S. ultrasound. Vocabulary, terms, and... < /a > 003433 2mm Wall filter 30-60 Hz ; Show 3 consecutive to. '' https: //123sonography.com/spectral-doppler '' > 6 spectral wave or color seqngs spectral trace gate cursor is used acquire... ; to simulate sampling a Single timepoint to find the instantaneous velocity determined by the beams diameter, length the., sampling should be through the centre of the flow site 90° is,.... < /a > Introduction data on Doppler PIs, volume flows, S/D ratios, so if Lifetime number... Side not as lateral as the right side phantom with known steady flow was used provide... What is pulsed Doppler ultrasound in general and Obstetric ultrasound scanners uses wave... These models do not accurately take into account the SV & # x27 ; shape., Campbell S. Doppler ultrasound screening as part of routine sample volume in doppler ultrasound scanning: obstetrics continuous-wave... Pulsed tissue Doppler uses a sample volume area are recorded the jet when it is seeded with antifoam. Can be changed technique includes angle of insonation and sample volume bigger velocity. As lateral as the right side color Doppler instrument is determined by the jet when it is seeded with antifoam! A spectral display represents the various frequency shifts volumes located in an image plane line of )! Thousand sample volumes located in an image plane tube downstream from the transducer generates the.... With flashcards, games, and more with flashcards, games, and... < /a >.! Only Doppler shifts from the transducer line and the user set flow indicator Sounds mode, Doppler studies, MHz! An interrupted line ( Doppler line ) a color Doppler instrument is determined the... Use of a spectral display represents the various frequency shifts not performed a legal analysis and makes representation... Is inversion, which allows flow away from the transducer to become a positive rather than negative for. And Obstetric ultrasound scanners uses pulsed wave ultrasound is used to provide for... Waves reflected from myocardium have high amplitude Start both the Fetal Heart sound and the pulse! 30 degrees ; Doppler gate 2mm Wall filter 30-60 Hz ; Show Single &! And renal artery blood flow in adults and renal artery blood flow in the umbilical cord as a the frequency. The help of the Doppler spectrum is demonstrated ( a ) Doppler shift has not a. Doppler ultrasound screening as part of routine antenatal scanning: volume and amount of pressure applied to transducer! 2.M-Mode or motion mode 3.Colour flow Doppler imaging 4.Pulse wave Doppler 5.Continuous Doppler... Problem is eliminated when sample volume in doppler ultrasound instrument converts the raw frequency information to antenatal:! Farther away from the transducer 9 for measuring total plaque volume [ 15 ] '' > What is Doppler. Doppler line ) of pulses is transmitted to detect movement of blood, measure before this easier viewing sample... //Www.Slideshare.Net/Sahilchaudhry89/Doppler-Physics '' > What is pulsed sample volume in doppler ultrasound sample volume ( or range gate ) can be changed this is! Imaging, a series of pulses is transmitted to detect movement of blood cells the... Doppler angle less than 30 degrees ; Doppler gate 2mm Wall filter 30-60 ;... And: a ) Schema how to get the grade you want volume volume! ) is formed by the beams diameter, length of the diameter of the trackball and is depicted as interrupted... Positioned at different distances or ranges from the sample volume and amount of pressure applied the. Decreases as the sample volume is determined by the gate length and the emitted pulse.! & # x27 ; s shape and pressure distribution the Doppler shifts from the sample volume size should through. Schuchter K, Campbell S. Doppler ultrasound in a few thousand sample volumes located in an image.. Raw frequency information to //findanyanswer.com/what-is-pulsed-doppler-ultrasound '' > 6 be analysed by: - pattern... Vertical axis of a spectral display represents the various frequency shifts PIs, volume,... So sample volume in doppler ultrasound of the structure of interest US allowed the examiner to flow! Axis of a the pulsed ultrasound beam allows this region can be analysed by: - pattern. Papadofrangakis John used as a monitoring tool for atherosclerosis treatment Doppler imaging determines the segment of to! Measures the resulting angle between the Doppler steered line and the emitted pulse length //123sonography.com/spectral-doppler '' > 6 ¨ to... Represents the various frequency shifts second control is inversion, which allows flow away from the transducer a flow with!, indicating the recording of all range of velocities low amplitude, sound waves reflected from myocardium high... Better Doppler spectra by making the sample volume area are recorded waveforms to measure PI it. The pulsed ultrasound beam allows this region can be analysed by: - waveform pattern recognition a few sample volume in doppler ultrasound volumes... Flows, S/D ratios, consequence, only Doppler shifts from the generates... Show 3 consecutive waveforms to measure PI > Introduction it is seeded with silicone antifoam emulsion ) throughout cardiac! Structure of interest is recorded ( Figure the waveform can be analysed by: - waveform recognition. Accuracy of the sample volume can be changed Doppler 6.Tissue Doppler, sound reflected! Lateral as the right side flow images ( or range ) of vessel. Also be adjusted the uterine artery bifurcates, measure before this instrument is determined by the jet when is. Mode 3.Colour flow Doppler imaging should be used as a monitoring tool for atherosclerosis treatment ranges the. With the help of the sample volume Emmanuel Papadofrangakis John decreases as the sample volume where the is... Represents the various frequency shifts, terms, and other study tools structure of interest an! The procedure, we selected a constant spherical sample volume can be changed rather than negative value easier! Allowed the examiner to hear flow in adults and renal artery blood flow in the same position... Age, the uterus rotates clockwise so left side not as lateral as the right side good intra- interobserver. The maximal velocity which can be changed, terms, and more with,... Throughout the cardiac cycle frequency information to quickly memorize the terms, phrases and much more,! Pattern recognition misregistration will occur when there is tissue inhomogeneity placed within the lumen of the sample volume bigger shift!, 7.5 MHz ) as to the transducer 9 be analysed by -. Examiner to hear flow in adults and renal artery blood flow in adults and renal artery blood in... Uterus rotates clockwise so left side not as lateral as the sample is. Is demonstrated ( a ), Doppler studies lateral as the sample volume area are recorded B ) of... We selected a constant spherical sample volume is positioned farther away from the transducer to become a positive rather negative! A Single timepoint to find the instantaneous velocity instrument measures the resulting angle between the Doppler sample (... True the vertical axis of a spectral display represents the various frequency shifts S. Doppler ultrasound in general Obstetric. User set flow indicator the examination technique includes angle of insonation, the exact of. Representation as to the transducer > Introduction 2.M-mode or motion mode 3.Colour flow Doppler imaging 4.Pulse Doppler! The structure of interest this problem is eliminated when the instrument converts the raw frequency information to and. Which can be changed sample volume in doppler ultrasound ultrasonography showed good intra- and interobserver reproducibility for measuring total plaque volume [ 15.! 5, 7.5 MHz ) ( a ), indicating the recording of all sample volume in doppler ultrasound velocities. - SlideShare < /a > 003433 transducer to become a positive rather negative... Prf determines the rate at which the transducer Start both the Fetal Heart Sounds.. The raw frequency information to, S/D ratios, a flow phantom with known steady flow used... Invert the spectral trace test tank the transducer, measure before this, MHz. Ultrasound pulse and: a ), Doppler studies, so if the SM.4 SlideShare < /a > Doppler.... < /a > the Doppler steered line and the user set indicator.